Comparison of intraocular pressure measurements obtained by icare pro rebound tonometer, tomey ft-1000 noncontact tonometer, and goldmann applanation tonometer in healthy subjects

Show simple item record

dc.contributor.author Guler, Mete
dc.contributor.author Bilak, Şemsettin
dc.contributor.author Bilgin, Burak
dc.contributor.author ve diğerleri
dc.date.accessioned 2022-07-07T05:54:06Z
dc.date.available 2022-07-07T05:54:06Z
dc.date.issued 2015
dc.identifier.issn 1057-0829
dc.identifier.uri http://dspace.adiyaman.edu.tr:8080/xmlui/handle/20.500.12414/3363
dc.description.abstract Purpose:To compare intraocular pressure (IOP) measurements obtained by Icare PRO rebound tonometer (RT), Tomey FT-1000 noncontact tonometer (NCT), and Goldmann applanation tonometer (GAT) in healthy subjects, and to investigate the influence of central corneal thickness (CCT) on IOP measurements.Methods:A total of 132 eyes of 66 healthy subjects were included in the study. All IOP measurements were applied in a sitting position and always in the same order (RT, NCT, and GAT). CCT was measured by noncontact anterior segment optical coherence tomography.Results:The mean CCT was 523.531.2 m. The mean IOPs with GAT, NCT, and RT measurements were 14.56 +/- 2.77, 13.42 +/- 2.99, 14.18 +/- 2.55 mm Hg, respectively. A significantly good agreement was found between RT and GAT measurements (P=0.515). The mean difference between RT and GAT measurements was -0.38 mm Hg. An underestimation was found in NCT measurements compared with GAT ones as defined from paired comparisons (P=0.003). The mean difference between NCT and GAT measurements was -1.14 mm Hg. There was no significant difference between NCT and RT readings (P=0.069). The mean difference between NCT and RT measurements was -0.76 mm Hg. There was a significant correlation between GAT-NCT (r=0.740, P<0.001), GAT-RT (r=0.743, P<0.001), NCT-RT (r=0.729, P<0.001) measurement methods. The increment in measured IOP for a 10 m increase in CCT was 0.16, 0.33, and 0.18 mm Hg for the GAT, NCT, and RT, respectively.Conclusions:The IOP readings with the new Icare PRO RT showed good correlation with those obtained by GAT. Tomey FT-1000 NCT underestimated the IOP compared with GAT. tr
dc.language.iso en tr
dc.publisher Lippincott Williams & Wilkins tr
dc.subject Central corneal thickness tr
dc.subject Goldmann applanation tonometer tr
dc.subject Intraocular pressure tr
dc.subject noncontact tonometer tr
dc.subject Rebound tonometer tr
dc.title Comparison of intraocular pressure measurements obtained by icare pro rebound tonometer, tomey ft-1000 noncontact tonometer, and goldmann applanation tonometer in healthy subjects tr
dc.type Article tr
dc.contributor.authorID 0000-0003-2552-4735 tr
dc.contributor.department Adiyaman Univ, Sch Med, Dept Ophthalmol, TR-02040 Adiyaman, Turkey tr
dc.contributor.department Adiyaman Univ, Dept Ophthalmol, Educ & Res Hosp, Adiyaman, Turkey tr
dc.identifier.endpage 618 tr
dc.identifier.issue 8 tr
dc.identifier.startpage 613 tr
dc.identifier.volume 24 tr
dc.source.title Journal of Glaucoma tr